Skip to content

Indirect enzyme-linked immunosorbent assay for the quantitative estimation of lysergic acid diethylamide in urine

Sarah Kerrigan, Donald. E. Brooks

Clinical Chemistry May 1, 1998 DOI: 10.1093/clinchem/44.5.985 (opens in new tab)

Study at a glance

AI-extracted from the abstract
Characteristics Assay development and validation Peer reviewed
Topics LSD
Citations 6
Key points The new indirect ELISA for LSD in urine has a limit of detection of 8 ng/L, analytical recoveries of 98–106%, and improved precision compared to the commercial assay.

Abstract

AbstractA new antibody to lysergic acid diethylamide (LSD) was used to develop a novel indirect ELISA for the quantification of drug in urine. Evaluation of the new assay with the commercially available LSD ELISA (STC Diagnostics) shows improved performance. The test requires 50 μL of urine, which is used to measure concentrations of drug in the μg/L to ng/L range. The limit of detection was 8 ng/L compared with 85 ng/L in the commercial assay, and analytical recoveries were 98–106%. Our test detected 0.1 μg/L of LSD in urine with an intraassay CV of 2.4% (n = 8) compared with 6.0% for a 0.5 μg/L sample in the commercial assay (n = 20). The upper and lower limits of quantification were estimated to be 7 μg/L and 50 ng/L, respectively. Specificity was evaluated by measuring the extent of cross-reactivity with 24 related substances. Drug determination using the new assay offers both improved sensitivity and precision compared with existing methods, thus facilitating the preliminary quantitative estimation of LSD in urine at lower concentrations with a greater degree of certainty.