Mindscape Collective is now The Consciousness Library. Same library, new name. You may need to sign in again. About the change
Skip to content

Detection of Lysergic Acid Diethylamide (LSD) in Urine by Gas Chromatography-Ion Trap Tandem Mass Spectrometry

J. H. Sklerov, Kathryn S. Kalasinsky, C. A. Ehorn

Journal of Analytical Toxicology October 1, 1999 DOI: 10.1093/jat/23.6.474 (opens in new tab) via OpenAlex

Summary

AI-generated from the abstract

A confirmatory method using gas chromatography-tandem mass spectrometry (GC-MS-MS) with an internal ionization ion trap detector can detect and quantify lysergic acid diethylamide (LSD) in urine. After a single-step solid-phase extraction of 5 mL of urine, underivatized LSD is measured with limits of quantitation and detection of 80 and 20 pg/mL, respectively. Temperature-programmed on-column injections show linearity over 20-2000 pg/mL (r² = 0.999). Intraday and interday coefficients of variation are below 6% and 13%, respectively. The method has been applied to quality-control specimens and LSD-positive samples, with comparisons to alternative GC-MS methods and extraction procedures.

Study at a glance

Characteristics Method development and validation Peer reviewed
Population Urine samples (quality-control and LSD-positive)
Topics LSD
Keywords Urine Detection limit Tandem mass spectrometry Gas chromatography–mass spectrometry Ion trap
Citations 19
Key finding A GC-MS-MS method using an ion trap detector can detect LSD in urine at concentrations as low as 20 pg/mL with high precision and linearity.

Abstract

A confirmatory method for the detection and quantitation of lysergic acid diethylamide (LSD) is presented. The method employs gas chromatography-tandem mass spectrometry (GC-MS-MS) using an internal ionization ion trap detector for sensitive MS-MS-in-time measurements of LSD extracted from urine. Following a single-step solid-phase extraction of 5 mL of urine, underivatized LSD can be measured with limits of quantitation and detection of 80 and 20 pg/mL, respectively. Temperature-programmed on-column injections of urine extracts were linear over the concentration range 20-2000 pg/mL (r2 = 0.999). Intraday and interday coefficients of variation were < 6% and < 13%, respectively. This procedure has been applied to quality-control specimens and LSD-positive samples in this laboratory. Comparisons with alternate GC-MS methods and extraction procedures are discussed.