Application of the Syva EMIT and Abbott TDx Amphetamine Immunoassays to the Detection of 3,4-Methylenedioxymethamphetamine (MDMA) and 3,4-Methylenedioxyethamphetamine (MDEA) in Urine

Gary W. Kunsman, Joseph E. Manno, K.r. Cockerham, Barbara R. Manno

Journal of Analytical Toxicology May 1, 1990 DOI: 10.1093/jat/14.3.149 (opens in new tab) via OpenAlex

Summary

AI-generated from the abstract

MDMA and MDEA, hallucinogenic amphetamine analogs, are increasingly used recreationally, prompting evaluation of common urine drug tests. The Abbott TDx and Syva EMIT-d.a.u. amphetamine immunoassays were tested for cross-reactivity with MDMA- and MDEA-spiked urine samples. The Syva EMIT assay detected only the highest concentration (10.0 micrograms/mL) of both substances. The TDx assay showed cross-reactivity for MDMA ranging from 18% at high concentrations to 118% at low concentrations, and for MDEA from 12% to 47%. Precision for TDx was good, with coefficients of variation below 6% for MDMA. Negative EMIT but positive TDx results were confirmed by GC/MS.

Study at a glance

Characteristics Laboratory evaluation Peer reviewed
Population MDMA- and MDEA-spiked urine samples
Topics MDMA
Keywords Chemistry Amphetamine Urine Chromatography
Citations 34
Key finding The Abbott TDx amphetamine assay showed substantial cross-reactivity with MDMA (18-118%) and MDEA (12-47%), while the Syva EMIT-d.a.u. assay detected only the highest concentration of both substances.

Abstract

MDMA and MDEA are hallucinogenic analogs of amphetamine. The need for laboratory monitoring of these substances has developed as a result of their increased recreational use. Since the Abbott TDx and Syva EMIT-d.a.u. immunoassays are commonly used tests for urine monitoring of drugs-of-abuse, the amphetamine assay of each manufacturer was assessed to determine the degree of cross-reactivity with MDMA and MDEA. Cross-reactivity was evaluated using a series of MDMA- and MDEA-spiked urine samples. Testing was performed over a two-day period with 3 runs/day and each sample run in duplicate. The Syva EMIT d.a.u. amphetamine assay was positive only at the highest concentration standard (10.0 micrograms/mL) for both MDMA and MDEA. Consequently, no further testing was performed with this assay. Calibration curves were generated for the TDx runs and percent cross-reactivity determinations were made. Precision for the TDx data was evaluated based on within-day and between-day coefficients of variation (CV). CVs for MDMA runs were below 6% for within-day and below 5% for between-day runs. Values of CV for MDEA were below 16% for both within-day and between runs; CVs were less than 2.5% for positive values. Cross-reactivity for MDMA ranged from 18% (10.00 micrograms/mL) to 118% (0.15 microgram/mL). Cross-reactivity for the MDEA standards ranged from 12% (10.00 micrograms/mL) to 47% (0.15 micrograms/mL). The presence of MDMA and/or MDEA in samples resulting in a negative EMIT-d.a.u. test and a positive TDx test was confirmed by GC/MS analysis.(ABSTRACT TRUNCATED AT 250 WORDS)